       USE 
        Prepared for use in a breath alcohol simulator which is capable of controlling 
        the solution temperature at 34 ± 0.2°C. When air (breath) is bubbled through 
        this solution, an equivalent BAC is generated based upon the 2100: 1 Breath 
        - Blood Alcohol Ratio. 
       PREPARATION 
         
        Absolute ethyl alcohol (USP Grade) is mixed with distilled water to provide 
        an analytical concentration predictive of the equivalent BAC as noted 
        above. For example, a 100 mg% (0.10%) would have an analytical concentration 
        of 1.21 mg/mL. A tolerance of 0.00 to +0.02 mg/mL is maintained.  
      ANALYSIS 
        Each batch of solution is independently analyzed by gas chromatography 
        using the internal standard technique. The certification process also 
        includes a simulator test using a Breathalyzer model 900/900A and an Intoxilyzer 
        5000. A certificate is available upon request.  
     | 
     
       BOTTLING 
         
        The solution is bottled in high-density polyethylene bottles and closed 
        with a poly seal cap. A neck seal is provided for integrity.  
      CONCENTRATION 
        Ethyl alcohol standard solutions are available in standard concentrations 
        to provide alcohol breath instruments with the following readings: 0.04% 
        BAC, 0.05% BAC, 0.10% BAC, and 0.15% BAC. Other concentrations are available. 
         
      LABELING 
        Each batch of standard solution is given a unique lot number and is further 
        identified by the equivalent BAC value (e.g. 0.10% BAC), the date of manufacture 
        and the expiry date for use. Each bottle and each box of six (6) bottles 
        is so labeled.  
     |